Surface and Subsurface Analyses of Metal-on-Polyethylene Total Hip Replacement Retrievals

dc.contributor.authorVuong, Vicky
dc.contributor.authorPettersson, Maria
dc.contributor.authorPersson, Cecilia
dc.contributor.authorLarsson, Sune
dc.contributor.authorGrandfield, Kathryn
dc.contributor.authorEngqvist, Håkan
dc.contributor.departmentMaterials Science and Engineeringen_US
dc.date.accessioned2022-05-10T15:36:09Z
dc.date.available2022-05-10T15:36:09Z
dc.date.issued2016-05
dc.description.abstractMetal-on-polyethylene (MoP) articulations are one of the most reliable implanted hip prostheses. Unfortunately, long-term failure remains an obstacle to the service life. There is a lack of higher resolution research investigating the metallic surface component of MoP hip implants. This study investigates the surface and subsurface features of metallic cobalt chromium molybdenum alloy (CoCrMo) femoral head components from failed MoP retrievals. Unused prostheses were used for comparison to differentiate between wear-induced defects and imperfections incurred during implant manufacturing. The predominant scratch morphology observed on the non-implanted references was shallow and linear, whereas the scratches on the retrievals consisted of largely nonlinear, irregular scratches of varying depth (up to 150 nm in retrievals and up to 60 nm in reference samples). Characteristic hard phases were observed on the surface and subsurface material of the cast samples. Across all samples, a 100-400 nm thick nanocrystalline layer was visible in the immediate subsurface microstructure. Although observation of the nanocrystalline layer has been reported in metal-on-metal articulations, its presence in MoP retrievals and unimplanted prostheses has not been extensively examined. The results suggest that manufacturing-induced surface and subsurface microstructural features are present in MoP hip prostheses prior to implantation and naturally, these imperfections may influence the in vivo wear processes after implantation.en_US
dc.identifier10.1007/s10439-015-1462-9
dc.identifier.citationVuong, V., Pettersson, M., Persson, C. et al. Surface and Subsurface Analyses of Metal-on-Polyethylene Total Hip Replacement Retrievals. Ann Biomed Eng 44, 1685–1697 (2016). https://doi.org/10.1007/s10439-015-1462-9en_US
dc.identifier.issn10.1007/s10439-015-1462-9
dc.identifier.urihttp://hdl.handle.net/11375/27541
dc.language.isoenen_US
dc.publisherSpringeren_US
dc.subjecttribologyen_US
dc.subjecttotal hip joint replacementen_US
dc.subjectelectron microscopyen_US
dc.titleSurface and Subsurface Analyses of Metal-on-Polyethylene Total Hip Replacement Retrievalsen_US
dc.typePostprinten_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Vuong-ABME-2016.pdf
Size:
733.67 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.68 KB
Format:
Item-specific license agreed upon to submission
Description: