Welcome to the upgraded MacSphere! We're putting the finishing touches on it; if you notice anything amiss, email macsphere@mcmaster.ca

Implant resonance and the mechanostat theory: Applications of therapeutic ultrasound for porous metallic scaffolds

dc.contributor.authorDeering, Joseph
dc.contributor.authorPresas, Alexandre
dc.contributor.authorYu, Bosco
dc.contributor.authorValentin, David
dc.contributor.authorHeiss, Christian
dc.contributor.authorBosbach, Wolfram A.
dc.contributor.authorGrandfield, Kathryn
dc.contributor.departmentMaterials Science and Engineeringen_US
dc.date.accessioned2021-07-13T19:31:11Z
dc.date.available2021-07-13T19:31:11Z
dc.date.issued2021-06
dc.description.abstractThe development of treatment strategies for improving secondary stability at the bone-implant interface is a challenge. Porous implants are one solution for improving long-term implant stability, but the osteoconduction process of implants into the bone can be slow. Strain-driven osteogenesis from the mechanostat theory offers insight into pathways for post-operative treatment but mechanisms to deliver strain to the bone-implant interface need refinement. In this work, the use of therapeutic ultrasound is simulated to induce resonance into a porous implant structure. Local strains through the scaffold are measured by varying systemic variables such as damping ratio, applied vibrational force, primary bone-implant stability, and input frequency. At the natural frequency of the system with applied forces of 0.5 N and a damping ratio of 0.5%, roughly half of the nodes in the simulated environment exceed the microstrain threshold of 1000 με required for new bone formation. A high degree of sensitivity was noted upon changing input frequency, with minor sensitivities arising from damping ratio and applied vibrational force. These findings suggest that the application of therapeutic resonance to improve osseointegration of the bone-implant interface may be viable for applications including dental implants or segmental bone defects.en_US
dc.identifier10.1016/j.msec.2021.112070
dc.identifier.citationDeering J, Presas A, Yu B, Valentin D, Heiss C, Bosbach WA, et al. Implant resonance and the mechanostat theory: Applications of therapeutic ultrasound for porous metallic scaffolds. Mater Sci Eng C. 2021;125:112070.en_US
dc.identifier.issn10.1016/j.msec.2021.112070
dc.identifier.urihttp://hdl.handle.net/11375/26667
dc.language.isoenen_US
dc.publisherElsevieren_US
dc.subjectMechanostat theoryen_US
dc.subjectOsseointegrationen_US
dc.subjectResonanceen_US
dc.subjectOsteogenesisen_US
dc.subjectPorous implantsen_US
dc.subjectImplant designen_US
dc.titleImplant resonance and the mechanostat theory: Applications of therapeutic ultrasound for porous metallic scaffoldsen_US
dc.typePostprinten_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2021_MSEC_Deering_Grandfield.pdf
Size:
1.24 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.68 KB
Format:
Item-specific license agreed upon to submission
Description: