Welcome to the upgraded MacSphere! We're putting the finishing touches on it; if you notice anything amiss, email macsphere@mcmaster.ca

Assessment of Various Thermoplastic Masks Used for Radiation Therapy Immobilization in Terms of Shrinkage and Pressure

dc.contributor.advisorOstapiak, Orest
dc.contributor.authorTunkr, Ayman
dc.contributor.departmentMedical Physicsen_US
dc.date.accessioned2014-12-02T20:11:31Z
dc.date.available2014-12-02T20:11:31Z
dc.date.issued2015-06
dc.description.abstractThe aims of this study are first, to assess and compare the amount of deformation for seven thermoplastic masks from three different manufacturers [Aquaplast RT-1889YS (Aqua variable perf), Fiberplast RT-1889KYS (Fiber variable perf), Embrace RT-8320 (Embrace), AccuPerf IMRT head mask RG461-4SHT (AccuPerf), Fiberplast RT-1889KS (Fiberplast), Aquaplast RT-1989S (Aqua thick), and Aquaplast RT-1889S( Aquaplast)] using time lapse photography, and second, to measure the amount of pressure exerted by each mask on a rigid head phantom using pressure sensitive film. For the first aim, all masks were marked with fiducial points located near forehead, nose, and chin. The masks were fabricated on a rigid head phantom following manufacturer instructions. Each mask was mounted on a plastic mounting tray, and photographed in a sagittal view over the span of one week. The fiducial coordinates were determined by a center of mass algorithm. A scale factor was determined by photographing a graph sheet mounted in place of the mask. For the second aim, the pressure exerted by each mask was evaluated by mounting the mask on the head phantom after sandwiching a strip of pressure sensitive film between the two at three anatomical locations: the forehead, nose, and chin. Measurement of translation in the posterior direction did not give reproducible results as direct caliper measurements did not agree with those from the time-lapse photography technique. In the superior/inferior direction, the Embrace mask showed the greatest deformation at forehead and nose areas (1.069±0.017mm (inferiorly), and 0.432±0.010mm (inferiorly) respectively), while the Aqua thick mask showed the greatest deformation at chin area (0.970±0.016mm (superiorly)). The averaged median pressure of the masks at the forehead location were not significantly different except for the Fiber variable perf which showed the greatest averaged median pressure ( 0.118 MPa). At the nose location, no mask showed a significantly different averaged median pressure. At the chin location, the Aqua thick mask showed a significantly higher averaged median pressure compared to the others (0.180 MPa).en_US
dc.description.degreeMaster of Applied Science (MASc)en_US
dc.description.degreetypeThesisen_US
dc.identifier.urihttp://hdl.handle.net/11375/16526
dc.language.isoenen_US
dc.titleAssessment of Various Thermoplastic Masks Used for Radiation Therapy Immobilization in Terms of Shrinkage and Pressureen_US
dc.typeThesisen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
master thesis Ayman .pdf
Size:
1.39 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.68 KB
Format:
Item-specific license agreed upon to submission
Description: