Title: Simultaneous correction for scatter and attentuation in positron emission tomography using statistical image reconstruction methods
Authors: Manji, Nekmohamed
Advisor: Nahmias, Claude
Department: Physics and Astronomy
Keywords: Astrophysics and Astronomy;Physics;Astrophysics and Astronomy
Publication Date: 2001
Abstract: <p>In this work, a method to perform simultaneous scatter and attenuation correction for PET is developed within the context of statistical image reconstruction methods. The principles are demonstrated using both numerical simulations of artificially constructed scatter distributions and Monte Carlo studies of phantoms, showing the viability of this method in practice. In addition, a new transmission scan protocol is defined which is necessary to collect the required data in order to perform the corrections. The method has several distinct advantages over traditional filtered back-projection techniques. These include treating scatter and attenuation simultaneously so that a correction can be made during image reconstruction, leading to a one-step method; guaranteed non-negativity of the reconstructed pixel intensity values; predictable noise propagation; and improved counting performance by 65% resulting in an increased signal-to-noise ratio (on the order of 28%), achieved using non-windowed acquisition. Noise Equivalent Count rate analysis indicates that for a nominal clinical activity concentration of 37 kBq/cc (1 μCi/cc), a three-fold improvement in scanner performance can be realized for a standard acquisition window, compared to an uncorrected image and a five-fold improvement is achievable using an extended acquisition window.</p>
URI: http://hdl.handle.net/11375/7279
Identifier: opendissertations/2563
3637
1396069
Appears in Collections:Open Access Dissertations and Theses

Files in This Item:
File SizeFormat 
fulltext.pdf
Open Access
8.04 MBAdobe PDFView/Open
Show full item record Statistics


Items in MacSphere are protected by copyright, with all rights reserved, unless otherwise indicated.

Sherman Centre for Digital Scholarship     McMaster University Libraries
©2022 McMaster University, 1280 Main Street West, Hamilton, Ontario L8S 4L8 | 905-525-9140 | Contact Us | Terms of Use & Privacy Policy | Feedback

Report Accessibility Issue